Cross calibration of QDR-2000 and QDR-1000 dual-energy X-ray densitometers for bone mineral and soft-tissue measurements

  • B. Abrahamsen*
  • , J. Gram
  • , T. B. Hansen
  • , H. Beck-Nielsen
  • *Corresponding author for this work

    Research output: Contribution to journalArticleResearchpeer-review

    Abstract

    Replacement of dual energy X-ray densitometry (DXA) equipment may be necessary in many labs with time, or new equipment may be added. In this context we compared patient scans between a Hologic QDR-1000W and a QDR-2000 (n = 29-43, depending on anatomic region) and between QDR-2000 single beam (SB) and fan beam (FB) modes (n = 40-62) as a quality control measure. A total of 83 subjects (79 females and four males) with a wide range of bone mineral densities (BMD) were studied. There was a linear relationship between results with the QDR-1000W and QDR-2000 in SB mode, and between SB and FB mode on the QDR-2000, but the magnitude of the coefficients and constants differed for the different anatomic regions. In SB mode the QDR-2000 underestimated whole body and forearm BMD by 3% relative to the QDR-1000W, even after cross calibration using a spine phantom. Femoral total BMD was slightly, but not significantly, underestimated. Thus, additional postanalysis correction had to be applied to forearm and whole-body scans. In FB with the QDR-2000, spine and whole-body BMD was underestimated by 3%, but femur total and neck BMD was overestimated by 2.2 and 2.8%, respectively, compared with SB scans on the same device. Soft-tissue composition with FB (enhanced analysis protocol) on the QDR-2000 differed greatly from that obtained using SB (standard protocol). Lean tissue mass was 4 kg lower and fat mass 4 kg higher in FB mode. Also, total body BMC was 7% lower using FB. Patients should be followed using the same scanning mode and analysis protocol whenever possible. Cross calibration by means of a single BMD anthropomorphic spine phantom is not sufficient to ensure identical results in all anatomic regions when a QDR-2000 replaces a QDR-1000W or is switched to FB mode. Linear regressions in vivo should be determined for each anatomic region.

    Original languageEnglish
    Pages (from-to)385-390
    Number of pages6
    JournalBone
    Volume16
    Issue number3
    DOIs
    Publication statusPublished - Mar 1995

    Funding

    Acknowledgments: This investigation is part of the Danish Osteoporosis Prevention Study supported by the Karen Elise Jensen Foundation. The excellent assistance of technicians Else Rettmer and Kirsten Westermann is gratefully acknowledged.

    Keywords

    • Bone
    • Bone mineral density
    • Densitometry
    • Dual-energy X-ray absorptiometry
    • Longitudinal studies
    • Precision