    Aims: The conventional analysis of dobutamine-atropine stress echocardiography (DASE) is poorly defined and subject to considerable variation. The aim of this study was to investigate the reproducibility of strictly standardised qualitative analysis in DASE. Methods and results: Strict criteria for standardised DASE interpretation were defined through logistic regression analysis on categorical parameters obtained from 20 patients with coronary artery disease (CAD) and 20 healthy controls subjected to DASE. Three expert echocardiographers analysed DASE recordings from 100 consecutive patients referred for coronary angiography. Specificity for CAD and for predicting significant stenosis of a major coronary artery was 94% (95% CI: 83-100%) and 79% (95% CI: 63-96%), whereas sensitivity was 49% in both cases (95% CI: 38-60% and 37-61%). Within and between observer reproducibility was moderate to fair (κ = 0.56 and 0.38; 95% CI: 0.40-0.72 and 0.24-0.52). In patients without prior myocardial infarction and in echogenic patients within observer reproducibility was good (κ = 0.72 and 0.74; 95% CI: 0.52-0.92 and 0.56-0.92). Conclusions: Observer variation was not eliminated in standardised qualitative DASE interpretation based on criteria that predicted the presence of CAD with a high specificity and reproducibility was good only in certain subgroups of patients.

    Original languageEnglish
    Pages (from-to)169-179
    Number of pages11
    JournalInternational Journal of Cardiac Imaging
    Volume18
    Issue number3
    DOIs
    Publication statusPublished - 2 Jul 2002

    Funding

    This study was funded by grants from: The Danish Heart Foundation, Copenhagen, Denmark, Eli Lilly Denmark A/S, Copenhagen, Denmark, Mc-Kinnys Foundation, Copenhagen, Denmark and GE Vingmed Ultrasound, Horten, Norway. For expert statistical assistance we thank the Biostatistical Unit, University of Copenhagen, Denmark.

    Keywords

    • Coronary artery disease
    • Diagnosis
    • Echocardiography
    • Stress testing

    Fingerprint

    Explore the research areas of 'Between observer variation is not eliminated by standardised analysis of dobutamine-atropine stress echocardiography'.

    Cite this