TY - JOUR
T1 - Mismatch 'never events' in hip and knee arthroplasty
T2 - a cohort and intervention study
AU - Odgaard, A
AU - Laursen, M B
AU - Gromov, K
AU - Troelsen, A
AU - Kristensen, P W
AU - Schrøder, H
AU - Madsen, F
AU - Overgaard, S
PY - 2019/8
Y1 - 2019/8
N2 - AIMS: The aim of this study was to give estimates of the incidence of component incompatibility in hip and knee arthroplasty and to test the effect of an online, real-time compatibility check.MATERIALS AND METHODS: Intraoperative barcode registration of arthroplasty implants was introduced in Denmark in 2013. We developed a compatibility database and, from May 2017, real-time compatibility checking was implemented and became part of the registration. We defined four classes of component incompatibility: A-I, A-II, B-I, and B-II, depending on an assessment of the level of risk to the patient (A/B), and on whether incompatibility was knowingly accepted (I/II).RESULTS: A total of 26 524 arthroplasties were analyzed. From 12 307 procedures that were undertaken before implementation of the compatibility check, 21 class A incompatibilities were identified (real- or high-risk combinations; 0.17%; 95% confidence interval (CI) 0.11 to 0.26). From 5692 hip and 6615 knee procedures prior to implementation of the compatibility check, we found rates of class A-I incompatibility (real- or high-risk combinations unknowingly inserted) of 0.14% (95% CI 0.06 to 0.28) and 0.17% (95% CI 0.08 to 0.30), respectively. From 14 217 procedures after the introduction of compatibility checking (7187 hips and 7030 knees), eight class A incompatibilities (0.06%; 95% CI 0.02 to 0.11) were identified. This difference was statistically significant (p = 0.008).CONCLUSION: Our data presents validated estimates of the baseline incidence of incompatibility events for hip and knee arthroplasty procedures and shows that a significant reduction in class A incompatibility events is possible using a web-based recording system. Cite this article: Bone Joint J 2019;101-B:960-969.
AB - AIMS: The aim of this study was to give estimates of the incidence of component incompatibility in hip and knee arthroplasty and to test the effect of an online, real-time compatibility check.MATERIALS AND METHODS: Intraoperative barcode registration of arthroplasty implants was introduced in Denmark in 2013. We developed a compatibility database and, from May 2017, real-time compatibility checking was implemented and became part of the registration. We defined four classes of component incompatibility: A-I, A-II, B-I, and B-II, depending on an assessment of the level of risk to the patient (A/B), and on whether incompatibility was knowingly accepted (I/II).RESULTS: A total of 26 524 arthroplasties were analyzed. From 12 307 procedures that were undertaken before implementation of the compatibility check, 21 class A incompatibilities were identified (real- or high-risk combinations; 0.17%; 95% confidence interval (CI) 0.11 to 0.26). From 5692 hip and 6615 knee procedures prior to implementation of the compatibility check, we found rates of class A-I incompatibility (real- or high-risk combinations unknowingly inserted) of 0.14% (95% CI 0.06 to 0.28) and 0.17% (95% CI 0.08 to 0.30), respectively. From 14 217 procedures after the introduction of compatibility checking (7187 hips and 7030 knees), eight class A incompatibilities (0.06%; 95% CI 0.02 to 0.11) were identified. This difference was statistically significant (p = 0.008).CONCLUSION: Our data presents validated estimates of the baseline incidence of incompatibility events for hip and knee arthroplasty procedures and shows that a significant reduction in class A incompatibility events is possible using a web-based recording system. Cite this article: Bone Joint J 2019;101-B:960-969.
KW - Arthroplasty, Replacement, Hip/instrumentation
KW - Arthroplasty, Replacement, Knee/instrumentation
KW - Cohort Studies
KW - Computer Systems
KW - Denmark
KW - Hip Prosthesis/adverse effects
KW - Humans
KW - Knee Prosthesis/adverse effects
KW - Medical Errors/adverse effects
KW - Prosthesis Design/adverse effects
KW - Prosthesis Failure/adverse effects
KW - Registries
U2 - 10.1302/0301-620X.101B8.BJJ-2018-1124.R1
DO - 10.1302/0301-620X.101B8.BJJ-2018-1124.R1
M3 - Article
C2 - 31362543
SN - 2049-4394
VL - 101-B
SP - 960
EP - 969
JO - Bone and Joint Journal
JF - Bone and Joint Journal
IS - 8
ER -