Abstract
Patients are not able to call for help during a generalized tonic-clonic epileptic seizure. Our objective was to develop a robust generic algorithm for automatic detection of tonic-clonic seizures, based on surface electromyography (sEMG) signals suitable for a portable device. Twenty-two seizures were analyzed from 11 consecutive patients. Our method is based on a high-pass filtering with a cutoff at 150 Hz, and monitoring a count of zero crossings with a hysteresis of ±50 μV . Based on data from one sEMG electrode (on the deltoid muscle), we achieved a sensitivity of 100% with a mean detection latency of 13.7 s, while the rate of false detection was limited to 1 false alarm per 24 h. The overall performance of the presented generic algorithm is adequate for clinical implementation.
| Originalsprog | Engelsk |
|---|---|
| Sider (fra-til) | 579-85 |
| Antal sider | 7 |
| Tidsskrift | IEEE Transactions on Biomedical Engineering |
| Vol/bind | 59 |
| Udgave nummer | 2 |
| DOI | |
| Status | Udgivet - feb. 2012 |
Fingeraftryk
Udforsk hvilke forskningsemner 'Automated algorithm for generalized tonic-clonic epileptic seizure onset detection based on sEMG zero-crossing rate' indeholder.Citationsformater
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver